Diagnostic Performance of the NasoCheck Comfort Rapid Antigen Test for SARS-CoV-2 Detection on Nasal Swabs Compared with RT-PCR during the 2021 COVID-19 Epidemic Wave in Guédiawaye (Dakar, Senegal)
Mba Elhadji Bambo Diakhaby *
Faculty of Health Sciences (UFR des Sciences de la Santé), Gaston Berger University, Saint-Louis, Senegal.
Omar Gueye
Bacteriology-Virology Laboratory, Dalal Jamm National Hospital Centre, Guédiawaye, Dakar, Senegal.
Modou Ndiaye
Bacteriology-Virology Laboratory, Dalal Jamm National Hospital Centre, Guédiawaye, Dakar, Senegal.
Dieynaba Barry
Bacteriology-Virology Laboratory, Dalal Jamm National Hospital Centre, Guédiawaye, Dakar, Senegal.
Ndeye Coumba Toure-Kane
Bacteriology-Virology Laboratory, Dalal Jamm National Hospital Centre, Guédiawaye, Dakar, Senegal and Institute for Health Research, Epidemiological Surveillance and Training (IRESSEF), Diamniadio, Senegal.
*Author to whom correspondence should be addressed.
Abstract
Background: Rapid antigen diagnostic tests provided a timely complement to molecular testing during the COVID-19 pandemic. This study assessed the diagnostic accuracy of the NasoCheck Comfort nasal antigen test against RT-PCR in Guédiawaye, Senegal.
Methods: A prospective cross-sectional study was conducted at Dalal Jamm National Hospital Centre between 1 July and 31 August 2021. Two respiratory specimens were obtained during the same visit from each of 200 consenting adults: a nasal swab for NasoCheck Comfort and a naso-oropharyngeal swab for Sansure RT-PCR. Diagnostic accuracy, Wilson 95% confidence intervals (CIs) and Cohen’s kappa were calculated.
Results: Two antigen tests were invalid (1.0%), leaving 198 analysable pairs; 78 (39.4%) were RT-PCR positive. There were 73 true-positive, 5 false-negative, 120 true-negative and no false-positive antigen results. Sensitivity was 93.6% (95% CI 85.9–97.2%), specificity 100.0% (96.9–100.0%), positive predictive value 100.0% (95.0–100.0%), negative predictive value 96.0% (91.0–98.3%) and overall accuracy 97.5% (94.2–98.9%). Cohen’s κ was 0.95. False-negative RT-PCR Ct values ranged from 24 to 28.
Conclusion: In this 2021 laboratory cohort, NasoCheck Comfort showed high agreement with RT-PCR, with sensitivity and specificity point estimates above the WHO minimum performance benchmarks for antigen tests. Negative antigen results warrant clinical assessment and molecular confirmation when infection is still suspected.
Keywords: SARS-CoV-2, COVID-19, rapid antigen diagnostic test, diagnostic accuracy, sensitivity, specificity, RT-PCR, Senegal